① Test instrument—fatigue test machine
② Test method
The time in which each valve achieves the target transvalvular pressure difference (aortic valve 100mmhg, mitral valve 120mmhg) in the test should account for more than 5% of its time in each open and close cycle, and such kind of cycles should account for more than 95% of the entire test cycles.
③ Test basis
Bioprosthetic valves should be capable of withstanding at least 200 million cycles and maintain intact structural properties in the condition of the transvalvular pressure difference (aortic valve 100mmhg, mitral valve 120mmhg), under normotensive conditions as specified in ISO 5840-2:2015.